Ideal for local injection with Motorized Stereotaxic Microinjector.

Easy set-up for injection by inputting a target time and volume while selecting a syringe from the built-in reference table of major syringes from Hamilton. Real time progress monitor displays the time and volume injected with flow rate. Easy syringe lock with a simple latch mechanism.

IMS-20

* Stereotaxic instrument, stereotaxic micromanipulator and syringe are sold separately.

* IMS-20 is recommended to use with SMM series; robust micromanipulators.



IMS-20

Motorized Stereotaxic Microinjector

Hamilton's Syringe Reference Table

No. Inner Diameter Volume Hamilton's syringe code
1 0.103mm 0.5µl 7000.5
2 0.146mm 1µl 7001, 7101
3 0.206mm 2µl 7002, 7102
4 0.330mm 5µl 7105
5 0.343mm 5µl 75
6 0.461mm 10µl 1701
7 0.485mm 10µl 701
8 0.729mm 25µl 702, 1702
9 1.030mm 50µl 705, 1705
10 1.458mm 100µl 710, 1710
11 2.304mm 250µl 725, 1725
User Direct Input Direct Input -

* The parameters of Hamilton syringes shown above are preset in the reference table.

* Inner diameter and volume of syringe can be input directly for using other syringes not prepared in the reference table.


IMS-20

Motorized Stereotaxic Microinjector

Specification

Accessories included Power Cord (1.5m)
Connecting Cable (2m)
Drive source 5-phase Stepping Motor
Movement range 60mm
Working voltage AC100V ~ 240V, 50/60Hz
Power consumption 10W
Dimensions/Weight Drive unit W30 × D167 × H84mm, 426g
Control unit W180 × D95 × H260mm, 2.45kg

* Loadable syringes must be from ø6.5mm to less than ø9mm in outer diameter. Syringe is not available from Narishige.

* Footswitch IMS-20F is available for additional control. (Sold separately)


System Diagram

System Diagram

System Diagram


Black Circle refers to the model itself with standard attachments
White Circle refers to attachable main accessories